DKI Jakarta - Jakarta Pusat

DR6000

Harga :
Rp. 1 / unit
Merk :
HACH
Pembelian Minimal :
1 unit
Pengiriman :
3 hari sejak pembayaran
Lokasi :
DKI Jakarta - Jakarta Pusat
Dilihat :
3 Kali
Update Terakhir :
12-06-2019 11:24

Detail Produk

 

Spesifikasi DR6000

Alat Uji Kualitas Air

Your Water Testing Needs, All in One Spectrophotometer

The DR6000 has the most pre-programmed testing methods, including high-speed wavelength scanning across the UV and Visible Spectrum.

Accessories for High Volume and High Accuracy Testing Needs

A carousel sample changer allows up to seven sequential measurements. The Sipper Module, an instrument-controlled sample delivery system, increases precision by constant optical characteristics.

Advanced Quality Assurance at Your Fingertips

The DR6000 comes with integrated QA software for scheduling, documenting and interpreting all of your needed quality measurements.

Guided Procedures and Elimination of False Readings

The DR6000, when used with TNTplus® reagent vials, provides the accurate results you need by guiding you step-by-step through your testing procedures. With TNTplus, the instrument averages 10 readings and eliminates outliers, making scratched, flawed or dirty glassware a non-issue.

Automatically Avoids Errors

RFID* technology automatically updates the program calibration factors when you place a TNTplus reagent box near the DR6000. The instrument identifies chemistry expiration dates via a barcode on the vials, and detects chemistry coefficient factors to avoid errors that can occur in lot-to-lot variations in the chemistry. 
*RFID technology currently available only in US, Anguilla, American Samoa, Australia, Bolivia, Canada, Cayman Islands, Columbia, Dominican Republic, El Salvador, Federated States of Micronesia, Guam, Guatemala, Marshall Islands, New Zealand, Northern Mariana Islands, Palau, Panama, Puerto Rico, and US Virgin Islands.

 What's in the box?

Includes: printed instrument manual, power cords (115V and 230V), universal holder, dust cover, 1-in. square matched glass sample cells, and a universal-adapter

 Specifications

Data Logger:5000 data points (result, date, time, sample-ID, user-ID)

Dimensions (H x W x D):215 mm x 500 mm x 460 mm

Display:TFT 7 inch WVGA color touch

Enclosure Rating:IP20 with closed lid

Includes:1 x Power Cord (US, EU)
1x Universal-Adapter
1x Dust Cover
Matched pair of 1 inch glass sample cells
Printed multilingual basic user manaul (en, es, fr, ja, ko, pt, zh)

Interfaces:USB type A (2), USB type B, Ethernet, RFID module

Manual Languages:en, es, fr, ja, ko, pt, zh

Operating Conditions:10 - 40 °C, max. 80% relative humidity (non-condensing)

Operating Mode:Transmittance (%), absorbance and concentration (wavelength, time)

Optical System:Reference beam, spectral

Photometric Accuracy:5 mAbs @ 0.0 - 0.5 Abs
<1% @ 0.5 - 2.0 Abs @ 546 nm

Photometric Linearity:0.005 - 2 Abs
≤ 0.01 at > 2 Abs with neutral glass at 546 nm

Photometric Measuring Range:± 3 Abs

Power Requirements:100 - 240 V; 50/60 Hz

Preprogrammed Methods:> 240

Sample Cell Compatibility:Rectangular: 10, 20, 30, 50 mm, 1 inch; round: 13 mm, 16 mm, 1 inch

Optional 100 mm rectangular cell with additional adapter

Scanning Speed:900 nm/min (in 1 nm steps)

Specific Technology:RFID for easy method update, sample ID and Certificate of Analysis

Spectral Bandwidth:2 nm

Storage Conditions:-25 to 60 °C / max. 80% relative humidity (non-condensing)

Stray Light:KI-solution at 220 nm < 3.3 Abs/ < 0.05%

User Interface Languages:bg, cn, cz, da, en, es, fr, gr, hr, hu, it, jp, kr, nl, pl, pt, ro, ru, sl, sl, sv, tr

User Programs:200

Warranty:1 year

Wavelength Accuracy:± 1 nm

Wavelength Range:190 - 1100 nm

Wavelength Reproducibility:< 0.1 nm

Wavelength Resolution:0.1 nm

Wavelength Selection:Automatic, based on method selection

Weight:11 kg

Harga DR6000

 : CALL

Permintaan Penawaran