General Specifications |
Temperature |
|
15°C to 35°C, operating; -51°C to 71°C, non-operating (with internal battery pack switched off)
|
|
Relative Humidity |
|
<95% to 30°C, <75% to 35°C, non-condensing
|
|
Altitude |
|
<3049m (10,000 ft) operating, <12,195m (40,000 ft) non-operating
|
|
Vibration |
|
Per MIL-T-28800; Type III, Class 5, Style E
|
|
Safety |
|
Designed to IEC 348, 2nd edition; 1978 and ANSI/ISA-S82; and UL1244, 2nd edition 1980, CSA C22.2 No 231, and IEC 1010
|
|
Calibration documentation |
Reports of Calibration are optional |
Size |
|
--------
|
732B DC Standard:
|
13.4 cm H x 9.8 cm W x 40.6 cm D (5.28 in H x 3.86 in W x 15.98 in D)
|
|
--------
|
732B-7001 External Battery and Charger:
|
13.4 cm H x 9.8 cm W x 40.6 cm D (5.28 in H x 3.86 in W x 15.98 in D)
|
|
|
|
Weight |
|
-------- |
732B DC Standard: |
5.91 kg (13 lbs.) |
|
--------- |
732B-7001 External Battery and Charger: |
5.45 kg (12 lbs.) |
|
|
|
Power, battery life |
732B DC Standard: |
100V, 120V, 220V, 240V ac ± 10%; 50 Hz to 60 Hz; 66W max. Internal 12V lead acid, gelled electrolyte battery operates for 70 hours at 23°C when fully charged. Trickle-charged continually when external ac power is applied. |
|
-------- |
732B-7001 External Battery and Charger: |
100V, 120V, 220V, 240V ac ± 10%; 50 Hz to 60 Hz; 66W max. Internal 12V lead acid, gelled electrolyte battery operates for 70 hours at 23°C when fully charged. Trickle-charged continually when external ac power is applied. |
|